| CRT |
Type |
6-inch rectangular
type with internal graticule
0%, 10%, 90% and 100% markers. 8 x 10 DIV (1 DIV=1 cm) |
| Phosphor |
P31 |
| Accelerating Potential |
14kV |
| Illumination |
Continuously adjustable |
| Z-Axis Input |
Coupling: DC
Sensitivity: 5V or more
Maximum Input Voltage: 30V (DC+AC
peak) at 1kHz or less
Bandwidth: DC ~ 5MHz |
| Vertical
System |
Sensitivity |
2mV ~ 5V / DIV,11 step in 1-2-5 sequence |
| Sensitivity
Accuracy |
≤3% (5 DIV at the center
of display) |
| Vernier Vertical Sensitivity |
Continuously variable to 1 / 2.5 or
less panel-indicate value |
| Bandwidth (-3dB) |
DC ~ 200MHz (2mV/DIV : DC ~ 20MHz) |
| Rise Time |
1.75nS (2mV/DIV : 17.5 nS) |
| Signal Delay |
Leading edge can be monitored |
| Max. Input Voltage |
400V (DC+AC peak) at 1kHz or less |
| Input Coupling |
AC, DC, GND |
| Input Impedance |
1MΩ±2% // approx. 25pF |
| Vertical Mode |
CH1, CH2, DUAL (CHOP/ALT), ADD, CH2
INV |
| Bandwidth Limited |
20MHz |
| Common-Mode Rejection Ratio |
50 :1 or better at 50kHz |
| Dynamic Range |
8 div at 100MHz, 5 div at 200MHz |
| Horizontal
System |
Horizontal Modes |
MAIN (A), ALT, DELAY (B) |
| A (main) Sweep Time |
20nS ~ 0.5S / DIV, continuously variable
(UNCAL) |
| B (delay) Sweep Time |
20nS ~ 50mS / DIV |
| Accuracy |
±3% (±5% at X10 MAG) |
| Sweep Magnification |
X10 (maximum sweep time 2nS / DIV) |
| Hold Off Time |
Variable |
| Delay Time |
1µS ~ 5S |
| Delay Jitter |
Better than 1:20000 |
| Alternate
Separation |
Variable |
| Trigger |
Trigger Modes |
AUTO, NORM, TV |
| Trigger Source |
CH1, CH2, LINE, EXT, EXT/10 |
| Trigger Coupling |
AC, DC, HFR, LFR, NR |
| Trigger Slope |
"+" or "-"
polarity or TV sync polarity |
| Trigger Sensitivity |
| Mode |
Frequency |
INT |
EXT |
EXT/10 |
| AUTO |
10Hz~20MHz |
0.35 DIV |
50
mV |
500
mV |
| 20MHz~200MHz |
1.5
DIV |
150
mV |
1.5
V |
| NORM |
DC~20MHz |
0.35
DIV |
50
mV |
500
mV |
| 20MHz~200MHz |
1.5
DIV |
150
mV |
1.5
V |
| TV |
sync
signal |
1
DIV |
200
mVpp |
2
Vpp |
|
| Trigger Level
Range |
INT: ±4DIV or more
EXT: ±0.4V or more
EXT/10: ±4V or more |
| TV Triggering |
TV-V, TV-H, TV-LINE |
| TV-Line
Selection |
|
Standard
|
Filed
1
|
Filed
2
|
|
NTSC
(525H)
|
1H~263H
|
1H~262H
|
|
PAL
(625H)
|
1H~313H
|
1H~312H
|
|
SECAM
(625H)
|
1H~313H
|
1H~312H
|
|
| Max. External Input Voltage |
400V (DC+AC peak) at 1kHz |
| External Input Impedance |
1MΩ±5%, // approx. 25pF |
| X-Y
Operation |
Mode |
X-axis: selectable CH1, CH2, EXT,
EXT/10
Y-axis: selectable CH1, CH2, CH1 and
CH2 |
| Sensitivity Accuracy |
2mV ~ 5V/DIV ±3%
EXT: 0.1V/DIV ±5%
EXT/10: 1V/DIV ±5% |
| X-axis Bandwidth |
DC ~ 500kHz (-3dB) |
| Phase Error |
3º or less from DC ~ 50kHz |
| Output
Signal |
Trigger Signal Output |
Voltage: approx. 25mV/DIV into 50Ω
Frequency response: DC ~ 10MHz |
| Calibrator Output |
1kHz Squarewave, 2Vpp±2% |
| Cursor
Readout Function |
Cursor Measurement Function |
ΔV, ΔV%, ΔVdB, ΔT,
1/ΔT, ΔT%, Δθ |
| Cursor Resolution |
1 / 100 DIV |
| Effective Cursor Range |
Vertical: ±3DIV
Horizontal: ±4DIV |
| Panel Setting Display |
Vertical: V/DIV(CH1,CH2),
UNCAL,
ALT/CHOP/ADD, INV, probe factor, AC/DC/GND
Horizontal: S/DIV (MTB,DTB), UNCAL, X10MAG, delay time, HO
Trigger: source, coupling, slope,
level, TV-V, TV-H
Others: X-Y, lock, save/recall MEM 0-9 |
| Auto
Measurement Function |
Parameter Function |
FREQ, PERIOD, ±WIDTH, ±DUTY
( + or - polarity selected by trigger slope) |
| Display Digits |
Max. 6-digits, decimal |
| Frequency Range |
50Hz - 200MHz |
| Accuracy |
1kHz - 200MHz: ±0.01%
50Hz - 1kHz: ±0.05% |
| Measuring Sensitivity |
>2 div (Measuring source selected
from CH1 and CH2 as synchronous signal sources) |
| Special
Function |
Auto Set |
Input Channel: CH1, CH2
Frequency Response:
50Hz~50MHz |
| Panel Setting Save &
Recall |
10 sets |
| Panel Setups Lock |
Provided |
| Power Source |
AC 100V / 120V / 230V±10%,
50 / 60Hz |
| Accessories |
Power cord, Instruction
manual, Probes(10:1/1:1) |
| Size |
12.2"W X 5.9"H
X 18.5"D |
| Weight |
19.8 lbs. |